V Operation and Indication Functions
7 Registering Data for Optical Members
- 54 -
(4) Dichroic Mirror
You can register optional data for up to nine dichroic mirrors.
If an attached dichroic mirror is not listed in the presets, register the dichroic mirror.
1
Select “Edit” using the W key or the X key and then press
the [
] key.
EDIT FILTER DATA
Location:[Other1]
Name :[ABC ]
EX :[EX365/10 ] Edit
DM :[DM400 ] Edit
BA :[――――――――――] Edit
2
Select one location out of nine locations to register the
optional data using the PREV or NEXT key.
EDIT DM DATA
Location:[Other1]
Name :[ ]
3
Move the cursor to “Name” using the S key or the T key .
Using the numerical/alphabetical keys and the
W key or
the
X key, type in the wavelength of the dichroic mirror.
You can type in up to 10 characters.
EDIT DM DATA
Location:[Other1]
Name :[DM410
]
4
To continue registration of optional data, move the cursor
to “Location” using the
S key or the T key and repeat
steps 2 and 3 above.
EDIT DM DATA
Location:[Other2]
Name :[ ]
5
Press the MODE key to end the setting and to return to the
OPERATION screen.
To delete registered data
1) Move the cursor to “Location” using the
S key or the T key and select data to delete using PREV or NEXT
key.
2) Press CLR key.