System does not turn on.
Probable Cause:
Wall outlet main power switch is in
O
(off) position
Suggestion:
Place switch in I (on) position.
Probable Cause:
System is not plugged in.
Suggestion:
Plug in system. Verify power cable is well seated in wall outlet
and in main power receptacle.
Probable Cause:
Wall circuit breaker is in off/tripped position.
Suggestion:
Turn on wall circuit breaker.
Probable Cause:
Key is absent or in OFF position.
Suggestion:
Insert key and rotate to ON position.
Probable Cause:
Internal system error.
Suggestion:
Turn key switch to OFF position, wait at least one minute and
then turn to ON position. If system fails to start, contact
service.