System User Interface is blank for more than 30 seconds.
Probable Cause:
Internal system error during boot-up.
Suggestion:
Press and hold power button to repower the tablet
Probable Cause:
Tablet battery is dead.
Suggestion:
Recharge tablet battery or operate with recharging cable
connected and plugged into wall outlet
No aiming beam is present when in READY mode and/or no laser
treatment light is delivered when the footswitch is depressed and/or the
beams are of poor quality.
Probable Cause:
Laser is in STANDBY mode, not READY mode.
Suggestion:
Select READY mode on control panel.
Probable Cause:
Footswitch is not connected.
Suggestion:
Connect footswitch. Check footswitch cable integrity.
Probable Cause:
Aiming beam is on low intensity setting.
Suggestion:
Adjust aiming beam intensity on user interface.
Probable Cause:
After five minutes of non-use, system goes to STANDBY.
Suggestion:
Switch mode from STANDBY to READY.
Probable Cause:
Remote interlock has been activated and has disabled
system.
Suggestion:
Ensure the action that activated remote interlock has ceased
and proceed.
Probable Cause:
Footswitch and/or footswitch cable damaged.
Suggestion:
Inspect for damage.