System User Interface is blank for more than 30 seconds.
Internal system error during boot-up.
Press and hold power button to repower the tablet
Recharge tablet battery or operate with recharging cable
connected and plugged into wall outlet
No aiming beam is present when in READY mode and/or no laser
treatment light is delivered when the footswitch is depressed and/or the
beams are of poor quality.
Laser is in STANDBY mode, not READY mode.
Select READY mode on control panel.
Footswitch is not connected.
Connect footswitch. Check footswitch cable integrity.
Aiming beam is on low intensity setting.
Adjust aiming beam intensity on user interface.
After five minutes of non-use, system goes to STANDBY.
Switch mode from STANDBY to READY.
Remote interlock has been activated and has disabled
system.
Ensure the action that activated remote interlock has ceased
and proceed.
Footswitch and/or footswitch cable damaged.