EasyManua.ls Logo

NOVA BIOMEDICAL BIOPROFILE FLEX2 - Page 20

NOVA BIOMEDICAL BIOPROFILE FLEX2
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-6
BioProfile FLEX2 ESM Instructions for Use Manual
3.2.3.1 Biosafety Cabinet Preparation (ambr® 15)
An existing sterile hood or biosafety cabinet may need preparation prior to installation 
of ambr® 15 and subsequent integration with FLEX2 + ESM, unlike ambr® 250 which is 
installed as a standalone biosafety cabinet. If an access opening on the left-side wall of
the hood does not already exist, one must be drilled in order to route the ESM External 
Tubing Set (CF, CW, and PC lines) from the sample cup mounted on the ambr® 15 deck 
out to the FLEX2 + ESM.
Care should be taken to position the access hole so that it creates the shortest distance
for the Sample (CF) line. As a general guideline, both the height and the depth of the 
access hole should be halfway between the inlet port on the right side of the FLEX2 and
the sample cup on the ambr® 15 deck. Use the diagram below as an additional reference.
If desired, a split gland-nut assembly supplied by Nova (PN 60251) may be purchased 
separately to seal the access hole. The CF, CW, and PC lines can be routed together 
through a single gland-nut assembly.
NOTE: It is the responsibility of the End User to prepare the FLEX2 + ESM bench
space and Biosafety Cabinet access hole(s) prior to ESM installation.
Figure 3.6 Left Side Wall of BioSafety Cabinet

Related product manuals