OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 12
Test Control:
The on-board control panel consists of a display screen, a Scroll Wheel,
and an Exit button.
The Scroll Wheel turns clockwise and counterclockwise. Turning the wheel
scrolls through menu and parameter options. Pushing the Scroll Wheel in
signals acceptance.
The Exit button cancels any pending input and returns to the Main Menu.
Refer to page 31 for more information about the onboard control panel.
Scroll Wheel Exit Button
Thermocouple:
The Consistometer uses two thermocouples for reading temperature. One is
inside the unit cabinet and measures the temperature of the oil bath around
the slurry cup.
The other is outside the unit cabinet. This one must be inserted into the top
of the cell cap so that it reaches down into the slurry cup and measures the
temperature of the slurry.
When the thermocouple is not in use, store it in the port on the right-hand
side of the cabinet to protect it from damage.