6. Turn the Main power switch on.
7. Set the date and time. See page 37.
8. To ll the oil reservoir, open the front cabinet door and remove the oil
reservoir cap. Pour approximately four liters (or until full) of oil into the
reservoir. Replace the cap. Make sure the seal is air tight. Use the sight
glass on the side of the reservoir to check the oil level.
Mineral Oil (#120-001) is recommended for tests up to 400°F. For
tests above 400°F, use only Phenylmethlsiloxane Silicone Fluid
(#120-004).
9. Turn the “Main Power” switch On.
10. Set the time and date on the on-board controller if they are not already
set. See page 37 for instructions.
11. Calibrate the potentiometer before connecting the Consistometer to a
computer. See page 38.
Reservoir