Place a beaker underneath the balance (not supplied)
and suspend a precision calibrated thermometer 0 °C to
30 °C on the edge of the beaker.
Record and store the temperature in the Setup Menu.
Tare the balance.
Press Enter button to start the Density application.
Follow directions of the prompts
“Place weight in Air, Press Enter”. Place the solid sample
on the pan and press the Enter button.
“Place weight in Liquid, Press Enter”. Suspend the solid
sample from the Weigh Below Hook into the beaker. Use
a fine wire or thread. Fill the beaker with auxiliary liquid
(liquid of known density
Q
0
' usually distilled water or
ethanol). Insure that the liquid will cover the sample by
at least 1 cm after immersion.Ensure that no air bubbles
adhere to the immersed part of the supporting wire. Re-
move air bubbles by moving the wire or by means of a
fine brush.
Weigh the solid (buoyancy
P
) by pressing the Enter
button.
The display indicates the density in grams/ cc.
A typical final display is shown which indicates all of the
parameters and values.
DEN
S
ITY 01/01/03 03:05:45 PM
TEMP
S
ET
U
P
S
TA
R
T
0g
6100g
*
Temp: 25.0 Air: 100.19 g
Aux D: 0.9971
L
iquid: 95.17 g
Volume: 5.03 g
19.90
g/cc