5.2 Specifications
Admissible Ambient Conditions
Use only in closed rooms
Ambient Temperature range: 5 ºC to 40 ºC
Atmospheric humidity: 80% rh @ to 30 ºC
Voltage fluctuations: –15% +10%
Installation category: II
Pollution degree: 2
Power supply voltage: 12 VAC, 50/60 Hz or 12 VDC, 1A
*** Units availability is country dependent.
Capacity (g) 12,000 22,000 32,0000
Readability (g) 0.1
Weighing units *** g, kg, lbs, oz, oz t, ct, dwt, taels (3), mommes, gn, ti, N, custom unit
Application modes Basic Weighing, Parts Counting, Differential Weighing, Statistics, Formulation,
Filling, Animal Weighing, Percent Weighing, Check Weighing, Density
Determination, Gross-Net-Tare Weighing, & SQC.
Features RS232 Port, GLP Protocol, Selectable Language, Display text, Selectable
Displayed Information Settings, Selectable Environment Settings, Selectable
Auto-Print Settings, Integral Weigh Below Hook, Contrast and Brightness
Control, Protective In-Use Cover
Repeatability (Std. dev.) (g) 0.1
Linearity (g) (+) 0.4
Tare range Full capacity by subtraction
Safe overload capacity 150 % of capacity
Stabilization time
<4 seconds
Sensitivity drift PPM/°C 3
(10 °C - 30 °C)
Operating temperature range:
w/internal calibration 10° to 40°C/50° to 104°F
w/o internal calibration 10° to 30°C/50° to 86°F
Power requirements External Adapter, 100-120 V ac, 220 V ac, 50/60 Hz
Plug configuration for US, Euro, UK, Japan & Australia
Display (in/cm) 0.6/1.5
Pan size (in/cm) 11 x 1 3/4 x 13/27.9 x 4.4 x 33
Dimensions
(WxHxD) (in/cm) 14 x 5 1/4 x 16/35.5 x 13.3 x 40.6