ESG-400
Error
no.
Error message Possible cause Remedial action
[…]
E645
Error
„Auto-restart”
„Contact OLY”
Internal software error. Send back to Olympus Service
Mains voltage drop
(only visible in error log)
„Auto-restart”
„Contact OLY”
Mains voltage too low. Check the power cord and the connection
to the wall outlet. Check the condition of
the power network of the facility.
Internal hardware error. Send back to Olympus Service
[…]
E649
Error
„Auto-restart”
„Contact OLY”
Internal hardware or software error. Send back to Olympus Service
E651
Error
„Auto-restart”
„Contact OLY”
Malfunction of Olympus HF
instrument.
Replace the instrument.
Internal hardware or software error. Send back to Olympus Service
Error
„Auto-restart”
„Contact OLY”
Internal software error. Send back to Olympus Service
Error
„Contact OLY”
Internal software error. Send back to Olympus Service
[…]
E656
Error
„Auto-restart”
„Contact OLY”
Internal software error. Send back to Olympus Service
Short circuit
Re-grasp tissue and
proceed.
Avoid having the jaws in
contact with each other.
„Contact OLY”
Electrodes of the HF instrument
may touch each other.
Ensure that the electrodes of the HF
instrument do not touch each other.
Malfunction of the HF instrument
and / or the connection cable.
Replace the HF instrument and / or the
connection cable.
Short circuit
Short circuit / Re-adjust
loop and proceed.
Avoid contact between
loop and other
instruments or metal
parts.
„Contact OLY”
Electrodes of the HF instrument
may touch each other.
Ensure that the electrodes of the HF
instrument do not touch each other.
Malfunction of the HF instrument
and / or the connection cable.
Replace the HF instrument and / or the
connection cable.
Error
„Auto-restart”
„Contact OLY”
Internal software error. Send back to Olympus Service
7.022.211 / ISSUE 5 71 / 110 Chapter 4: Troubleshooting