ESG-400
Error
no.
Error message Possible cause Remedial action
Short circuit
Re-grasp tissue and
proceed. Avoid contact
between the instrument
tip and grasper.
„Replace Instrument”
„Contact OLY”
Electrodes of the HF instrument
may touch each other.
Ensure that the electrodes of the HF
instrument do not touch each other.
Malfunction of the HF instrument
and / or the connection cable.
Replace the HF instrument and / or the
connection cable.
Error
Application time limit
exceeded
„Replace Instrument”
„Contact OLY”
Cut and coag switches were
pressed simultaneously.
Release the switches and proceed to
work by pressing only one switch (cut or
coag).
Malfunction of the HF instrument
and / or the connection cable.
Replace the HF instrument and / or the
connection cable.
[…]
E671
Error
„Auto-restart”
„Contact OLY”
Internal software error. Send back to Olympus Service
E673
Error
„Auto-restart”
„Contact OLY”
Internal software or hardware error. Send back to Olympus Service
7.022.211 / ISSUE 5 72 / 110 Chapter 4: Troubleshooting