89
STM7
Power supply unit MM6-ILPS-2
Refer to the instruction manual provided with the power supply unit MM6-ILPS-2.
Operating environment
Ambient temperature/
humidity
Accuracy guarantee Temperature: 20 °C±1 °C Humidity: 65%±20%
Operation guarantee Temperature: 10 °C to 35 °C Humidity: 30% to 85%
Standby condition Temperature: 0 °C to 40 °C Humidity: 20% to 85%
Storage condition Temperature: -10 °C to 60 °C Humidity: 10% to 90%
Conditions for safety
standards
· Indoor use.
· Altitude: Max. 2,000 meters
· Ambient temperature: 5 to 40 °C (41 to 104 °F)
· Maximum relative humidity: 80% for temperatures up to 31 °C (88 °F) (without condensation)
In case of over 31 °C (88 °F), the relative humidity is decreased linearly through 70% at 34 °C (93 °F),
60% at 37 °C (99 °F), and to 50% at 40 °C (104 °F).
· Supply voltage fluctuations: Not to exceed ±10% of the normal voltage.
· Pollution degree: 2 (in accordance with IEC60664-1)
· Installation/Overvoltage category: II (in accordance with IEC60664-1)