Safety Accessories Supplementary Sheet
46 OMICRON
WARNING
Death or severe injury caused by high voltage or current possible.
Contact OMICRON technical support if the CP Strobe Warning Lamp is not
flashing during the test.
7.5 Disconnection
To disconnect the CP Strobe Warning Lamp from the main device:
1. Make sure that the main device used with the CP Strobe Warning Lamp is switched off.
2. Press the Emergency Stop button on the main device.
3. Disconnect the CP Strobe connection cable from the CP Strobe Warning Lamp and the main device.
4. Disconnect the Safety Connector Dongle from the CP Strobe connection cable.
5. Connect the Safety Connector Dongle to the SAFETY output connector of the main device.
6. The CP Strobe has now been removed and the main device has the required Safety Accessory
Dongle installed on the SAFETY output connector.
7.6 Cleaning
To clean the CP Strobe Warning Lamp, use a cloth dampened with isopropanol alcohol.
WARNING
Death or severe injury caused by high voltage or current possible.
Before cleaning the CP Strobe, always disconnect it from the main device.