OMNI 4000/7000 Operations and Maintenance Guide Rev F
3-22
OMNICONNECT® Operations
To set up the plenum pressure option, follow these instructions:
1.
In the Actions ribbon, click Configure.
2.
a. Expand the Prover function in
the Configure tree.
b. Select Configuration and
Setup.
c. Verify that a Pulse Interpolation
type prover is selected in the
Prover Type screen.
3.
a. Click on General Setup in the
Configure tree.
b. Select the Equipment List tab in
the screens on the right.
4.
a. In the Prover group on the lower
right-hand side of the screen,
check the Plenum Pressure
option.
b. Click Apply to save your change.