EasyManua.ls Logo

Omron E5AR - MANUAL 2

Omron E5AR - MANUAL 2
343 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9.7 Inspecting indicator accuracy
9-13
User calibration
9.7 Inspecting indicator accuracy
After calibrating input, be sure to inspect the indicator accuracy to verify that the input was calibrated
correctly.
Run the E5AR/ER in the PV/SP state.
Check the upper limit, lower limit and mid-range limit of the indicator range (3 values).
Thermocouple • Preparations
Connect as follows to the required devices. Be sure to connect the
E5AR/ER to the cold junction compensator using the compensation
wire that you intend to use for the thermocouple.
• Operation
Make sure that the cold junction compensator is at 0°C, and set the
STV output to the voltage that is equivalent to the inspection value
startup power.
If the cold junction compensating system is set to external, a cold
junction compensator and compensation wire are not needed.
Resistance
temperature input
sensor
• Preparations
Connect as follows to the required devices.
• Operation
Set the 6-dial to the resistance that is equivalent to the inspection
value.
E5AR/E5ER
5
6
Cold junction compensator
STV
Compensating wire
_
+
Input
power
supply
1
A
2
Input power supply
1
A
2
5
4
A
B
B
6
6 dial
E5AR/E5ER

Table of Contents

Related product manuals