plus
UM - EHY2000plus-R24
10
Installation notices
1. This unit should be kept away from heat registers, radiators, stoves or
other appliances that produce heat. Windy places or the vicinity of
windows should also be avoided.
2. For the best performance and safety, please place this unit in the
middle of the room. Make it possible for the operator to access the
device from any direction. Any walls and/or grounded surface have to
be minimum 1.5 meter away from the treatment bed, so that the
patient cannot reach any surfaces independent from the device, which
could cause an electric shock for them.
3. The EHY unit must be located in a suitable place, where the emergency
switches of the device are accessible for everybody in case of a
dangerous situation. This emergency switch cuts off the mains voltage of
EHY device itself, but this button does not affect the waterbed and the
PC.
4. Good air circulation around the device is essential to prevent internal
overheat of the electronic parts.
5. Choose a dust-free, well-ventilated position. Do the everyday cleaning
by washing up the floor, as it is usual in medical institutions.
6. Take care of the intactness of cables and water pipes. Do not break
them, or obstruct the free flow of the cooling water.
7. Electrical safety of the appliance is only guaranteed if the grounding
system of the building is in accordance with the local electricity board
regulations. The OncoTherm service team controls it at the regular
services. The proper electric conditions are the liability of the customer.
8. Only authorized service personnel should carry out repairs and any
other work on the device. The approval must be in written form from the
Oncotherm Group. The relocation and/or using mains-socket other than
the originally installed, must be done only by the Oncotherm or its
authorized service-representative. Only the qualified and authorised
Oncotherm personnel should service the unit when it does not operate
normally or shows marked change in its performance.
9. Devices, which are to be discarded, must be made unusable. Pull out
the plug and remove the cable.