Installation
General
The device must be installed by a qualified technician/engineer on behalf of the
Oncotherm Group in compliance with the instructions provided. The
manufacturer declines all responsibility for improper installation which may harm
people and damage property.
When the packing is removed, check that the appliance is not damaged. If you
have any doubts, do not use the appliance, but call for a qualified technician.
The packaging items (plastic bags, foamed polystyrene, nails, etc.) are potential
sources of danger. Never leave them within the reach of children.
This device shall be used for the purpose it was expressly designed for. Any other
use is considered improper and consequently dangerous. The manufacturer
declines all responsibility for damage resulting from improper and irresponsible
use.
Electrical connection
1. Connect the equipment to 230/110V A/C socket with ground only.
Ensure that the socket is properly installed.
2. Observe that a min. 16 Ampere fuse protects the socket.
3. Make sure that the device uses the single-phase independently from
other appliances. Use an independent phase for the device and others
for the other applied electric appliances (for example: air-conditioning,
diagnostics systems, computers, sterilization equipment, etc.) in the
room.
(Our service team controls these conditions, and we will be taken into
consideration during installation to satisfy these requests.)