UM - EHY2000plus-R24
34
If the tuning is not successful within 90 seconds, the device will indicate it with a
sound signal and the “Error” LED will be blinking until the tuning is not
successful. In this case the electrode should be re-positioned to reach a better
electrical contact with the patient.
To pause the treatment for any reason (for example: to give an
injection) press the STOP button for a short period of time. Then you can
continue the treatment by pressing the START button again.
If you have finished the treatment, stop the device by pressing the STOP
button for a longer period of time.
Note: Some patients may need some relaxation after the treatment.
Switch off the device
Before you switch off the device it is suggested to finish the treatment
completely (stop it and do not pause it). After the complete stop, please wait
a few minutes before the complete shut - down to make it possible for the
internal cooling to cool down the important electronic parts in order to make
their lifetime longer.
After this, switch off the device by the mains switch.