Safety
1. The device is only suitable for normal treatment use and for the
purposes and intended uses stated in the operating instructions.
2. Do not use any extension- and radio-frequency-cables, only those
which are provided by the authorized service and/or by the
Oncotherm.
3. Before starting any cleaning work on the device it must be
disconnected from the electric supply by removing the plug from the
socket. Do not pull the cable!
4. The mains lead of the unit should be unplugged when the unit is not in
use.
5. Do not plug in or unplug the mains lead with wet hands.
6. Do not use the device when you are barefoot.
7. Make sure that objects do not fall and liquids are not spilled into the
interior of the device through openings.
8. Do not allow children to operate and/or control the equipment.
9. Do not allow people who are not experienced nor trained in operating
the equipment to use it without supervision.
10. Never leave the device exposed to environmental effects (rain, sun
etc.).
11. Every six months the service team has to check the device according to
the ‘regular safety testing instruction’.
12. Never use the waterbed without the mattress. Touching the heating
surface directly can cause burn!
13. Never use the mattress without water (only distilled water can be used.)
as the patient can compress it and can touch heating surface almost
directly! (Check of filling up: Push the middle of one half of the mattress
with one hand till reaching bottom. Then push the middle of the other
half with the other hand - minimum ½ meter from the first point. It should
not reach the bottom.)
14. Water leakage is shown on display (sensed electronically in wastewater
tank). In case of water-loss failure, treatment should be stopped
immediately and the service team should be called for help. If water
flows quickly, treatment must be stopped, switch off the bed and pull
out the mains plug.