UM - EHY2000plus-R24
69
Appendix 5 - Electromagnetic
compatibility information
According to IEC60601-1-2:2001 standard requirements
EHY-2000plus
EHY-2000plus device needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC information provided
below.
Portable and mobile RF communications equipment can affect EHY-2000plus
device.
Table 201 Guidance and manufacturer´s declaration electromagnetic
emission for all EQUIPMENT and SYSTEMS (see 6.8.3.201 a) 3))
Guidance and manufacturer’s declaration-electromagnetic emissions (T.201)
The EHY-2000plus device is intended for use in the electromagnetic environment specified below.
The customer or the user of the EHY-2000plus should assure that it is used in such environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 2
The EHY-2000plus must emit electromagnetic energy in order
to perform its intended function. Nearby electronic
equipment may be affected.
RF emissions
CISPR 11
Class B
The EHY-2000plus is suitable for use in all establishments,
including domestic establishments, and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/ flicker
emissions
IEC 61000-3-3
Complies