15. In case of noticing water in the plastic tray or on the surface of the
mattress, please check the water level in mattress and presence of
possible mechanical damage. Do not use damaged mattress in order
to avoid an electric shock.
16. The patient’s surface must not be damp. After cleaning the surface the
user has to wait until it is dry.
17. Using EHY-2000plus without a heated waterbed may decrease the
temperature of the body of the patient. In case of everyday use, keep
the bed switched on permanently. (Heating up the water from room
temperature to treating temperature level requires about 12 -16 hours.)
18. Check the temperature settings before each treatment!
Recommended temperature setting is 30-32°C, because the treatment
itself also heats and the patient will have sweaty skin which is not
suitable for correct EHY-2000plus treatment.
19. Check the alarm system daily during continuous use. Switch the device
off for a short time. When you switch on the waterbed, a self-test is
implemented, and a warning is shown until the user sets the desired
temperature again.
20. Do not use other surgical or endocardial devices while the patient is on
the waterbed.
Resident risks:
The user has to follow the instructions mentioned above, otherwise burning or
overheating of the tissue can occur.