Important notices
Choosing the “Important notices” link in the “Main menu” you can read the
most important notices and warnings about the use of the device.
Important notes for the use of the software
The following notes should be taken into consideration and followed when
using the web box.
• The treatment manager view can be used only on a computer from
where the patient is within sight.
• The patient should be monitored continuously not only through the web
server but in the room as well.
• All warnings mentioned in the EHY-2000plus user’s manual have to be
considered
• The logged computers must be certified and linked on to the web server
by a service technician.
• The treatment view shows the actual time on every new loaded page. It
should be checked every time the page is really newly loaded or comes
from the cache.
• The device is matched with a colour (and a serial number). In case of
every control the web page colour and the serial number written in the
border should be checked if the right device is controlled.
• After entering power or time, please check in the treatment window, if
the parameter is really set and if is the parameter which is required.
• The RF radiation emitted by the device may jam the WLAN connections,
that is why it is prohibited to use WLAN to connect the device with the
controlling computer!
Supported browsers
The used protocol is the primary HTTP protocol with cookies. This means that
any nowadays used common browser should be able to view the pages
supported by the web processor. The suggested browser is Mozilla Firefox. If
Internet Explorer is used, security level of the browser should be set to medium
or low.
The following browsers have been tested with the web box:
1. Mozilla Firefox 5.0
2. Microsoft Internet Explorer 7.0