Display-WEY-2000plus waterbed
Control panel of waterbed
“Temp. fault” lamp: 1 yellow lamp shows if the temperature is not correct. It
alerts the user if the surface temperature differs with more than +/-1 ºC from
the setting in stabilized state.
“No sensor” lamp: 1 yellow lamp alerts the user if one or more sensors do not
function correctly. It can occur if the mattress is pulled out of its place.
“Water loss” lamp: It lights up if water flows into the wastewater tank (approx. 3
liters) from mattress. This tank is at the foot-end of the waterbed. The tank’s
capacity is 20 liters. If flowing is not stopped, water will appear around the
mattress. With wet mattress the treatment cannot be continued. It can be
dangerous for the patient. (Danger of burn)
“Heating power On/Off” button: To turn off and on waterbed heating. The
lamp is yellow when the operating power is on.
Temp fault lamp
No sensor lamp
Heating power
On/Off
Cooling power
lamp
Temperature
displays
Water loss lamp
Danger!
Overheat lamp
Cooling
temp/Sound
reset
Set temp
Up/Down
Cooler status
display