EasyManua.ls Logo

OncoTherm EHY-2000plus - Page 74

OncoTherm EHY-2000plus
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
plus
UM - EHY2000plus-R24
73
WEY-2000plus
Table 201 Guidance and manufacturer´s declaration electromagnetic
emission for all EQUIPMENT AND SYSTEMS (see 6.8.3.201 a) 3))
Guidance and manufacturer’s declaration-electromagnetic emissions (T.201)
The WEY-2000plus device is intended for use in the electromagnetic environment specified below.
The customer or the user of the WEY-2000plus should assure that it is used in such environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
The WEY-2000plus uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The WEY-2000plus is suitable for use in all establishments,
including domestic establishments, and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies