Pulsar
2
Service Manual
OPTIKON 2000
Cod. 111004SEN Rev.C
6-10
7.
ELECTRIC I.V. POLE (if available) Switch the PULSAR2 off, conne
ct the
electric I.V. pole cable to the rear panel connector. Switch the equipment (and the
cart) on. Verify that the PULSAR2 tests the pole moving it slightly up and down.
Once in the working program, verify that the pole height can be adjusted with the
arrow keys in the Irrigation window and that the actual position of the pole
appears in the relevant display.
bar MPa PSI
0,1 bar
0.01MPa
1,45 PSI
0,2 bar
0.02MPa
2,90 PSI
0,3 bar
0.03MPa
4,35 PSI
0,4 bar
0.04MPa
5,80 PSI
0,5 bar
0.05MPa
7,25 PSI
1 bar
0.1MPa
14,508 PSI
2 bar
0.2MPa
29 PSI
3 bar
0.3MPa
43,52 PSI
4 bar
0.4MPa
58,03 PSI
5 bar
0.5MPa
72,54 PSI