rapid nebuliser system – Questions about your device? Contact your service partner 13
en
E Cleaning Your Nebuliser Handset
Step E3 Add a few drops of clear liquid dish soap to warm
(about 37 °C or 98 °F) tap water in a clean tub or
bowl and soak all parts of the nebuliser handset
for 5 minutes.
Fig. E3
Step E4 Rinse all parts thoroughly under warm running
tap water and then inspect them to make sure they
are clean and not visibly damaged.
If parts do not appear to be clean, soak them for
another 5 minutes as described in Step E3.
Do not use dirty or damaged parts.
Fig. E4
Step E5 Shake the parts to remove excess water.
Air-dry parts on a clean, lint-free towel.
After the parts are completely dry, store them in a
clean bag or cover them with a clean, lint-free towel.
Reassemble just before treatment.
Fig. E5
CAUTION
The membrane of the aerosol head can be
damaged very easily.
- Do not attempt to clean the aerosol head with a
brush.
- Do not attempt to clean or disinfect the aerosol
head in a home dishwasher.
- Do not attempt to disinfect the aerosol head in
a microwave.