6.5.1 Start/Stop
The Start/Stop measurement method implements an automatic data com-
pression feature and requires no further settings.
In addition to the current measured value, the Min and Max values are also
saved.
The measurement is completed independent of the current storage rate at a
scanning rate of 1 ms.
If the predefined memory area is full, the data compression is activated. The
storage rate is doubled and every second measured value is deleted. In this
case, the Min and Max values are also compared and the highest and lowest
measured value are transferred to the Min and Max value respectively.
INFORMATION
The Min and Max values are not lost but are saved, even in the case of long-
term measurements.
1 Tap on the Start
button to start a measurement.
2 Tap the Stop
button to stop the measurement.
3 Alternatively, the Start/Stop measurement ends automatically after 7 days.
51
The Parker Service Master COMPACT V1.0/02/24
Operation