EasyManua.ls Logo

Peekel Instruments PICAS - 5 Technical Specifications

Default Icon
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Peekel Instruments User manual PICAS V2.6.1
Page 58 of 59
5 Technical Specifications
Carrier frequency inputs of the CA2CF-card
General
Typical accuracy class 0.1%
Bandwidth (-3 dB) 2000 Hz
Maximum cable length: 500m
Sensor connection 2-, 3-, 4-, or 6-wire configurations
Bridge supply (transformer-isolated)
Supply voltage 0,5... 5V (adjustable)
Voltage accuracy ± 0.05%
Frequency 5 kHz
Frequency accuracy ± 1%
Load 60..-..1000 0,1%
1000..- 3000 >0,1%
Internal bridge-completion ½- bridge und
¼- bridge 120 / 350
Measuring input (transformer-isolated)
Ranges (@5V excitation):
± 100 µV/V.... ± 1 V/V
Input Filter: (High pass) > 500 Hz
Max. Common Mode Voltage 200V
Common Mode Rejection (50 Hz) >120 dB
Serial Mode Rejection: >66 dB
Capacitive input overload max.7x range permissible
Special input filtering for noise reduction
Balance control
R-balance +/- 65 mV/V
C-Balance at 120 bridge up to 10 nF
Output
Full scale voltage +/- 10 V
Protection long-term short circuit allowed
Maximum capacitive load 10 nF
Maximum cable length 100 m (@100 pF/m)
Frequency (-3 dB) < 2000 Hz
Filter type 7-pole low pass Butter worth -42 dB/Octave
Analog inputs of CA4AI-card
General
Typical accuracy class 0.1%
Bandwidth (-3dB) 10 Hz
Sensor connections 2-, 3-, or 4-wire configuratieons
Sensor supply:
Voltage: 5V ± 0.1% (max. 50 mA)
(Maximum for 4 channels together is 100 mA)
Current: 1mA ± 5% (max. 7,5k)
Power for active sensors 24VDC (max 80mA)
(galvanic separated from inputs)

Table of Contents