EasyManua.ls Logo

PerkinElmer Lambda 25 - Optics; Abscissa

PerkinElmer Lambda 25
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Lambda 25, 35, 45 User’s Guide
42
Optics
Beam center height 15 mm above cell holder bottom
Beam cross-section
1 nm slit ca. 0.6 mm x 9 mm (width x height)
at focal point of sample and reference beam in
sample compartment
Optical pathlength in
sample compartment
121 mm
Grating (Monochromator)
Holographic concave grating with 1053 lines/mm
in the center
Radiation sources Prealigned deuterium and halogen lamps
Detector
Photodiodes
(One for the sample beam and one for the
reference beam)
Abscissa
Wavelength range
190 nm to 1100 nm;
0 nm for alignment purposes
Wavelength accuracy ±0.1 nm
Wavelength reproducibility ±0.1 nm
Spectral bandwidth 1 nm (fixed slit)
Lamp change
Automatically at 326 nm
(software selectable over the whole
wavelength range)
Scan speeds
7.5, 15, 30 , 60, 120, 240, 480, 960,
1920 and 2880 nm/min
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com