EasyManuals Logo

PerkinElmer Lambda 25 User Manual

PerkinElmer Lambda 25
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
Installation
59
Connecting to the Electrical Supply
WARNING
Electrical Hazard
To prevent potential injury to yourself and damage to the instrument, first
make the electrical connections between the instruments in the system
before connecting to the line power supply.
The spectrometer automatically adjusts to the correct operating voltage.
Before starting the instrument for the first time, make sure that the correct
fuse is fitted to your line power supply. If you have a different fuse, change
it for the correct one. Do not connect the spectrometer to the line power
supply if the wrong fuse is fitted.
1. Make sure that the correct fuses are fitted in the holder at the rear of the
spectrometer (see Changing Fuses on page 102).
Voltage Standard
100 V – 120 V 6.3 A slow-blow
210 V – 240 V 3.15 A slow-blow
NOTE: The module has two fuses.
2. Make sure that the plug fitted to the line power cord provided with the
spectrometer is suitable for your local electrical outlets.
3. After all connections have been made between the various components of the
system, make certain that all power switches are set to off, then connect the
line cords to the electrical power supply.
The power switch is located at the top right-hand rear of the spectrometer.
NOTE: To prevent interferences caused by earth loops when operating with ancillary
instruments (autosamplers, etc.), connect all components of the system to the same
phase of the electrical supply via a multisocket distributor.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer Lambda 25 and is the answer not in the manual?

PerkinElmer Lambda 25 Specifications

General IconGeneral
BrandPerkinElmer
ModelLambda 25
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals