Installation
55
Preparing the Working Area
For maximum stability and minimum maintenance observe the following
requirements when siting the instrument:
• A firm base free from vibration.
• Enough space around and underneath the instrument for efficient air
circulation.
• A constant temperature between 15 ºC and 35 ºC.
• Constant humidity between 20% and 80% relative humidity.
• An atmosphere free from dust and corrosive fumes.
• Keep out of direct sunlight.
Illumination with diffuse lighting is ideal.
• A suitable source of electrical power should be located in the vicinity of the
instrument.
Electrical power must be available at a proper earth-grounded 3-wire electrical
outlet.
• The standard sample compartment baseplates have drain holes in them to run
off spilled liquids to the benchtop underneath the instrument. If required, a
sheet of thick filter paper can be placed under the instrument.