Installation
55
Preparing the Working Area
For maximum stability and minimum maintenance observe the following
requirements when siting the instrument:
A firm base free from vibration.
Enough space around and underneath the instrument for efficient air
circulation.
A constant temperature between 15 ºC and 35 ºC.
Constant humidity between 20% and 80% relative humidity.
An atmosphere free from dust and corrosive fumes.
Keep out of direct sunlight.
Illumination with diffuse lighting is ideal.
A suitable source of electrical power should be located in the vicinity of the
instrument.
Electrical power must be available at a proper earth-grounded 3-wire electrical
outlet.
The standard sample compartment baseplates have drain holes in them to run
off spilled liquids to the benchtop underneath the instrument. If required, a
sheet of thick filter paper can be placed under the instrument.