7.3 Switch on GRANOLYSER Warm-up phase
Switch ON the
GRANOLYSER using the main switch (position I).
The operating system starts up.
The halogen lamp switches on.
The motor starts so that any remaining sample is transported out of the instrument.
A 10-minute warm-up phase starts.
Progress of the warm-up phase
Skip warm-up phase
During the warm-up phase,
a flashing lamp symbol appears in
the yellow status bar.
NOTICE
If the warm-
up phase is skipped, there can be deviations or fluctuations in the
measurement result. In the case of approved instruments, a measurement is only
possible after the warm-up phase.
7.4 Start menu product selection
Product selection
Scroll product selection to
left and right
Display time and date
Information, operating instructions
and messages are displayed in dialog
boxes.