Maintenance and cleaning
40 | 46
Granolyser / Granolyser HL operating instructions | Revision 3
8.3 Inspection interval and function test
Sub-assembly Interval in one-shift working
Normal function tests: w m ¼ y ½ y y
Main switch I/0 X
Touchscreen X
Firm seating of magnets at the back on the drawer X
Mains cable and connection X
Labels and warning notes in place and legible (by visual
inspection)
X
Check the correctness and firm fit of the entire machine X
Check all cables for damage and aging X
Check protective covers for defects, as well as for correct
and tight fit
X
Check all plug-in, screw-in and clamp connections for firm
fit, tighten if necessary
X
Electrical test according to VDE See DGUV regulation 3
DANGER
Damaged cables increase the risk of getting an electric shock!
Never touch a damaged cable (e.g. cable that has been cut into, insulation damaged
etc.)
Disconnect the cable from the electrical power supply immediately.
Replace the cable with a new one.