Revision 3 | Granolyser / Granolyser HL operating instructions
2 Safety
NOTICE
It is strictly prohibited to deactivate the safety devices or modify their mode of effect.
2.1 Built-in safety systems
The built-in safety systems must be checked at regular intervals with the corresponding test methods, see
the following table:
d = daily
w = weekly
m = monthly
¼ y = every three months
½ y = every six months
y = yearly
V = Visual inspection
F = Function test
M = Measurement
Mains disconnector for a portable instrument
The main switch I/0 is the mains disconnector, and also serves as the
EMERGENCY OFF function. It is located on the rear of the instrument.
Figure 1: Rear GRANOLYSER HL main switch Rear GRANOLYSER main switch
Mains cable connection – power connector (IEC 60320 C14)
The connection for the mains cable (IEC 60320 C14 power connector) is located on the rear of the
instrument below the main switch.