Granolyser / Granolyser HL operating instructions | Revision 3
8.1.1 Housing
Cleaning Rectification Interval
Filling funnel
With a clean, dry and lint-free cloth.
Use a moist cloth in case of heavy contamination.
d
Drawer
With a clean, dry and lint-free cloth.
Use a moist cloth in case of heavy contamination.
d
Pane of glass in the
measuring chamber
Carefully with the soft brush supplied with the instrument, or using
a moist eyeglasses cleaning cloth.
ar
Housing
With a clean, dry and lint-free cloth.
Use a moist cloth in case of heavy contamination.
w
Touchscreen
Carefully with a clean, dry and lint-free cloth or a moist screen
cleaning cloth.
ar
Pfeuffer GmbH recommends a complete cleaning or covering of the instrument before
a longer standstill, in order to preserve the operational readiness of the GRANOLYSER.
8.1.2 Variant HL – Volume measuring cell and weighing equipment
Inside the GRANOLYSER HL arise dust deposits and impurities.
Remove the drawer.
Lay the
GRANOLYSER HL carefully on the side.
Remove the impurities inside carefully with a soft brush.
You can also use a vacuum cleaner for heavy contamination. This must be used without contact to
prevent damage to the measuring cell and weighing cell.
CAUTION
Pollutants such as hazardous levels of Mycotoxins can be present in this dust!
There can be respiratory ailments and infections can emerge!
Wear a respiratory mask!
Place the GRANOLYSER HL again properly and slide in the drawer.