Connecting to the ECG Output or Marker Input
Installin
g
the Instrument
87
Connecting to the ECG Output or Marker Input
See the specifications for the ECG Output and for the Marker Input under Interfaces in the
Monitor Performance Specifications section of the Specifications chapter of the Instructions
for Use, and the documentation for the device you are connecting.
Configuring the Monitor
See the Basic Operation chapter and the Configuration chapter of the Instructions for Use, as
well as the setup sections for each of the measurements for more information on Monitor
setup.
There is a second level of configuration for the Monitor which is only available after entering
a password. See the Installation chapter and Configuration chapter of the User’s Guide for
more information.
Selecting NBP Measurement Characteristic
Only software revisions D and later allow the NBP Measurement Characteristic to be
changed.
WARNING Consult technical support before changing the NBP measurement characteristic in
Adult or Pediatric mode. Changing the NBP measurement characteristic will
significantly change the NBP readings, especially in the elevated pressure range.
The setting of NBP Measurement Characteristic is not transferred as part of a cloning
operation with the cloning tool. It has to be set on each monitor separately.
NOTE This setting changes the NBP measurement characteristic in adult and pediatric mode. It has
no influence to the NBP measurement characteristic in neonatal mode. In neonatal mode
always the intraarterial method is used as reference standard to validate the overall accuracy.
For the M3046A Monitor with software revision E:
Step 1 Highlight the NBP numeric and press on the TouchStrip.
OR
Step 1 Press hte Setup key.
Step 2 Move the highlight to “NBP”.
Step 3 Press on the TouchStrip.
(When you are finished with the NBP Setup, press the Main Screen key.)