Test Reporting
138
Testin
g
the Instrument
Test Reporting
The following table shows what must be recorded on the Service Record after completing the
tests in this chapter.
Where P = Pass, F = Fail and X/x are the measured values as defined in the tests described in
this chapter
Test What to record
Visual V:P or V:F
Power OnPO:P or PO:F
P NIBP PN:P/X1/X2/X3/X4 or
PN:F/X1/X2/X3/X4
P CO
2
PCO2:P/X1/X2/X3/X4/X5/X6/X7/X8 or
PCO2:F/X1/X2/X3/X4/X5/X6/X7/X8
Safety S(1):P/x1/x2 or
S(1):F/x1/x2
S(2): P/x1 or
S(2): F/x1
S(3): P/x1 or
S(3): F/x1