Monitor Theory of Operation
Introduction to the Instrument
15
IrDA/Serial Communication Manager
The IrDA/Serial Communication manager is responsible for sending raw data to the local
recorder (serial protocol) or printer (in a format that complies with the IrDA (Infrared Data
Association) standard). The IrDA/Serial Communication manager provides a general printer
or recorder device interface to the printer or recorder manager, and maps the general printer/
recorder services to the device protocol. The general printer services are as follows:
Establishing and terminating the printer/recorder connection.
Transferring data to the printer/recorder.
Retrieving the printer/recorder status.
The IrDA/Serial Communication manager feeds the printer/recorder status (time out, paper
out, and so on) back to the printer or recorder manager, which in turn notifies the user of any
errors in the print/record process.
Communication Software Module
The communication module maintains a data link between the Monitor, the Measurement
Server and the Measurement Server Extension, and controls the exchange of data between
them. This data includes the following:
Measurement data.
status information.
Control information.
Configuration data.
Support Services
The support services software module contains miscellaneous functions that both the
Monitor, the Measurement Server and the Measurement Server Extension require. These
functions are as follows:
Date and Time.
Settings Handler.
Status Revision and Display.
Heart Rate Selector.