Troubleshooting the Installed Instrument
Troubleshootin
the Instrument
213
Troubleshooting When There is No Message on the Screen
Symptom Possible Cause What To Do
Some or all of the
numerics or waves
are missing from
the screen.
No measurements
connected
Check that a Measurement
Server and all the required
transducers are connected.
Connect a Measurement
Server.
No transducers
connected
Connect the required
transducers.
Defective transducer Replace the suspect
transducer.
Measurement Server
defective
Exchange the Measurement
Server.
The Measurement Server
should be investigated
technically as soon as
possible.
Monitor screen
dim
Brightness controls not
properly adjusted
Adjust brightness controls.
The screen may not be as
bright when the Monitor is
operating from the battery.
Monitor screen
blank
Power is not connected
or switched on
Connect power and switch on
the Monitor.
Battery not installed or
empty (Battery LED
flashes red, or flashes
red when you press the
On-Off/Standby switch.)
Fit a charged battery and
switch on the Monitor.