82
3.2.5.3.
Trigger slope
Depress
Y/\
switch
SK4, AUTO
switch
SK10,
+
switch
SK8 and
BEAM
SELECTOR switch
A SK14
Set
TIM
E/cm switch
SK2 to
.2 ms and TIM E/cm
potentiometer
R2
to
CAL
Set
AMPL switch
SKI
7
to
.1 V/cm and AMPL
potentiometer
R4
to
CAL
~
Apply
a
sine-wave signal
of
600
mVp.p, frequency 2 kHz,
to
Y/\
input
socket
BUS
Adjust
LEVEL
potentiometer
R3
for
a
stationary display
Check that the display starts with the
positive
slope
of
the signal
Depress
switch
SK9
Check that
the
display
starts with the negative
slope of
the signal
3.2.5.4.
Level range with external triggering
Depress EXT switch
SK6, AC
switch
SK1
1,
+
switch
SK8
and
BEAM
SELECTOR
switch
A SKI
4
Set
AMPL switch SKI 7
to
10 V/cm and
AMPL
potentiometer R4
to
CAL
Set
TIME/cm switch
SK2 to .2 ms and
TIME/cm potentiometer
R2 to CAL
Apply
a
sine-wave
voltage of
20
Vp.p,
frequency
2
kHz,
to both TRIGG Input
socket
BU2 and Y/\
input
socket BUS
Check
that
the time-base generator
cuts
out when LEVEL potentiometer RS is turned
fully
anti-clockwise
and
fully clockwise
Depress
switch SK9 and repeat the
check
3.2.
5.5.
Level and level
range
-T-
Depress Y/\ switch
SK4, AUTO
switch SK10,
+
switch
SK8
and BEAM
SELECTOR
switch
A SK14
Set
AMPL
switch SKI 7 to .1 V/cm and
AMPL
potentiometer R4 to CAL
Set
TIME/cm
switch
SK2
to .2 ms and
TIME/cm potentiometer R2 to CAL
Apply a
sine-wave
voltage
of
600 mVp.p, frequency 2 kHz, to
Y^
input
socket
BUS
Check that
the starting point of the trace moves upwards when
LEVEL potentiometer
RS is turned
clockwise
Check that the
starting
point of
the
trace
can be shifted over 5 cm
by
means
of LEVEL
potentiometer RS
Depress AC switch SK1
1
Set AMPL switch SK17 to
50
mV/cm
Check that the time-base
generator
cuts
out
when LEVEL
potentiometer
RS
is
turned fully
anti-clockwise
and
fully clockwise
Set AMPL SK17 switch
to
20
mV/cm
Check
that the time-base
generator does not
cut out when
LEVEL
potentiometer
RS is turned
fully
anti-
clockwise
and fully
clockwise
3.
2.
5.6.
Mains triggering
-
Depress EXT
switch SK6,
+
switch
SK8, AUTO
switch
SK10 and BEAM
SELECTOR
switch
A SK14
Set TIME/cm switch SK2
to 20
ms and TIME/cm potentiometer R2
to
CAL
Apply a
voltage
with the mains
frequency
to
Y^
input socket BUS; trace height
1 cm
Check
that the
oscilloscope
displays
a
moving
sine
wave
Depress MAINS switch
SK7
Check
that
the display is
stationary

Table of Contents