3.2.10.
Z
MODULATION
Depress
switch SK4,
+
switch
SK8,
AUTO
switch
SK10
and
BEAM SELECTOR
switch
A SK14
Apply
a
positive-going square-wave
voltage
of
20 V, repetition
rate
100 Hz, to
Z MOD
socket
BU7
Apply
a
signal which
is synchronous
with
the
former
square-wave voltage
to
Y/\ Input
socket
BUS
Check
that
the trace
is blanked during
the positive pulse
3.2.11.
OUTPUT
SOCKETS
3.2.11.1.
Calibration voltage on
CAL
socket
BUI (R761)
Check,
with the
aid of an accurate peak-peak
voltmeter, that
the voltage on
CAL
socket
BUI has
a
value
of 600
mVp.p
+
or
0,7
%;
if necessary,
readjust potentiometer
R761 (Fig.
3.12.)
Check
that the pulse does
not
show
overshoot
and that the repetition frequency
is 2
kHz
+
or
20
%
3
2.11.2.
Time-base voltage on
TB OUT socket
BU6
Depress
AUTO switch
SK10
Set
TIME/cm switch
SK2 to
2
ms
Check
that the amplitude
of
the sawtooth voltage
on
TB OUT
socket
BUG is 6 V
+
or
20
%
86
Fig,
3,
12,
Adjusting
Fig,
3,
U.