7. Remove the two philips screws securing the Inlet Filter to the Spine.
8. Remove the Inlet filter from the Spine.
Installation: (for item identification, refer to removal instructions above)
1. Place the Inlet Filter into the cavity of the spine assembly.
2. Secure the Inlet Filter to the spine assembly by securing the two screws to 6 in.-lbs.
3. Connect the Inlet Hose to the Inlet Filter.
4. Install the Compressor.
5. Install the Compressor Cover.
6. Install the Sieve Bed.
7. Install the Front Enclosure.
8. Install the Rear Enclosure.
7.18 LOSS OF POWER ALARM REPLACEMENT
Removal:
1. Remove the Data Port Cover
2. Remove the Power Loss Alarm
Installation: (for item identification, refer to removal instructions above)
1. Install the Loss of Power Alarm
2. Install the Data Port Cover