PAGE 5-11097626, VER. 03
CHAPTER 5: ALARMS & TROUBLESHOOTING
5.0 CHAPTER OVERVIEW
This chapter identifies the alarms associated with the SimplyGo Portable Oxygen Concentrator and SimplyFlo
Transportable Oxygen Concentrator. This chapter should be used by service technicians to help diagnose
problems with the SimplyGo or SimplyFlo, along with determining what parts, if any, need to be replaced.
5.1 SIMPLYGO ALARM INDICATORS AND SCREEN SYMBOLS
VISUAL, AUDIO INDICATORS DESCRIPTION
3 short beeps + 1 long beep
No Breath Alarm
This alarm occurs when a breath is not detected for a period of 2
minutes or more. It can be silenced by pressing the Alarm symbol
shown on the screen. This alarm becomes silent as soon as a breath is
detected. If no breath is detected after approximately 30 minutes, the
device shuts down to conserve power. (This alarm is inactive for ten
minutes after startup.)
3 long beeps + 1 short beep
Low Oxygen Concentration Alarm
This alarm occurs when the device is delivering a lower concentration of
oxygen than specified. This alarm can be silenced by pressing the
Alarm symbol.
1 long beep after 2 minutes
High Breath Rate Alarm
This alarm indicates that the user’s breath rate is exceeding the capacity
of the device. The device is still working properly and is still providing
oxygen at the rate specified for the setting. The beeps can be silenced
by pressing the Alarm symbol.
3 short beeps + 3 long beeps
Technical Fault Alarm
This alarm occurs when the device has a general malfunction and the
device is no longer operating properly.
The device may or may not shut down depending on the severity of the
alarm.
This alarm can be silenced if the Alarm Silence Symbol appears.
1 short beep
Low Battery Alarm
This alarm occurs when approximately 17 minutes of battery life
remains. Remaining battery life is dependant on your device settings
and your activity level. The empty battery symbol flashes on-screen.