PAGE 5-71097626, VER. 03
5.3 SIMPLYGO/SIMPLYFLO DATA RECOVERY TOOL AND MANIFEST FILE
This Software is designed to recover and view events from the SimplyGo and SimplyFlo devices and also
provide “real-time” viewing of certain system parameters during operation. The Tool is to be used for system
monitoring and troubleshooting, if required. Readings displayed by the indicators in the real time viewer shall
not be used for measurements or readings required for final testing of the SimplyGo/SimplyFlo.
5.3.1 COMPUTER SYSTEM & EQUIPMENT REQUIREMENTS
• Computer with Microsoft Windows 7 (32/64 bit) or XP (32 bit) operating systems only.
• Oxygen Interface Hardware Kit, Respironics Part Number 1073499, includes the following:
• 1 ea. USB-MiniDIN7 Communication Module
• 1 ea. USB Interface Cable
• 1 ea. MiniDIN Interface Cable