PAGE 3-2
1097626, V
ER. 03
3.2 OXYGEN CONCENTRATOR
Oxygen
Concentration*
87% - 96% at all settings
SimplyGo Flow
Settings and Pulse
Volumes
Pulse Mode
1 = 12 ml; 1 1/2 = 18 ml; 2 = 24 ml; 2 1/2 = 30 ml; 3 = 36 ml;
3 1/2 = 42 ml; 4 = 48 ml; 4 1/2 = 54 ml; 5 = 60 ml; 5 1/2 = 66 ml;
and 6 = 72 ml;
+/- 15% or 4 ml, whichever is greater (Average of 20 consecutive
breaths)
up to a max of 2000 ml/min +/- 200 ml
Sleep Mode
Variable pulse volumes, based on breath rate, to maintain a
constant minute volume per setting.
1 = 250 ml; 1 1/2 = 375 ml; 2 = 500 ml; 2 1/2 = 625 ml;
3 = 750 ml; 3 1/2 = 875 ml; 4 = 1000 ml; 4 1/2 = 1125 ml;
5 = 1250 ml; 5 1/2 = 1375 ml; and 6 = 1500 ml
+/- 15% or 60 ml, whichever is greater (Average of 20 consecutive
breaths)
Continuous Mode
1/2 = 500 ml/min; 1 = 1000 ml/min; 1 1/2 = 1500 ml/min;
2 = 2000 ml/min;
+/- 10% or 100 ml/min, whichever is greater (3 minute running
average)
NOTE: Max recommended flow is 2 LPM (at nominal outlet
pressures of 0 and 7 kPA)
SimplyFlo
Continuous Flow
Settings and
Volumes
Continuous Mode
1/2 = 500 ml/min; 1 = 1000 ml/min; 1 1/2 = 1500 ml/min;
2 = 2000 ml/min;
+/- 15% or 150 ml/min, whichever is greater (3 minute running
average)
NOTE: Max recommended flow is 2 LPM (at nominal outlet
pressures of 0 and 7 kPA)
Dimensions 11.5 in x 10 in x 6 in
(29.2 cm x 25.4 cm x x15.2)
SimplyGo Weight 10.0 lbs (4.54 kg) with battery installed
SimplyFlo Weight <9.0 lbs (4.1 kg)
User Interface Push Buttons, back-lit liquid crystal display (LCD)
SimplyGo Sound
Level
43 dBA max. at Nominal Pulse Mode Setting of 2 and 20 BPM
SimplyFlo Sound
Level
50 dBA typical at 2 LPM continuous
Audio Alarm 50 decibels (nominal) at 39 in. (1 m)