PAGE 7-26
1097626, V
ER. 03
7.13 SIEVE CANISTER REPLACEMENT
Removal:
1. Remove the Rear Enclosure.
2. Remove the Front Enclosure.
3. Remove the four philips screws securing the Sieve Canister to the spine assembly.
4. Remove the Canister and two O-rings.
Installation: (for item identification, refer to removal instructions above)
1. Insert the two new O-rings into position on the canister.
2. Secure the Sieve Canister to the spine assembly by tightening the four screws.
3. Install the Front Enclosure.
4. Install the Rear Enclosure.
WARNING
The top of the sieve canister must point toward the Patient Filter. The top of the Sieve Canister can be
identified by having a spring under the clear plastic.