PAGE 5-191097626, VER. 03
5.3.6 SIMPLYGO/SIMPLYFLO TROUBLESHOOTING
PROBLEM / EVENT CODE POSSIBLE CAUSE RECOMMENDED ACTIONS
Device Won’t Turn On
SimplyGo Battery is
depleted
Use the AC or DC power cords to
operate the device (with the battery
inserted) to recharge the battery.
SimplyGo Battery is not
installed correctly.
Remove the battery and re-install it
correctly.
Power button not
pressed twice
Press the power button twice.
SimplyGo Will Not Trigger A
Pulse Of Oxygen
SimplyGo is not turned
on.
Press the power button twice.
Cannula tubing is
kinked or twisted.
Make sure the tubing is connected
properly to the oxygen outlet port and
that it is free of any obstruction.
Oxygen Not At Full
Concentration
Device is warming up Wait 10 minutes for the device to
deliver oxygen at the prescribed
concentration.
Low O2 Concentration
Leaks
Air Inlet Filter
Faulty Compressor
Defective process
valves
Faulty Sieve Canister
Check the spine using snoop leak
detector and replace if necessary.
Ensure all tubing is properly
connected.
Remove Inlet Filter, operate device. If
concentration increases replace filter.
Verify and Replace the compressor
Check valve connections or replace the
defective valve
Replace both sieve canisters if needed.
SimplyGo High Breath Rate
Exceeded
User caused event Normal event - no action necessary
Low O2 Sensor failure
Main PCA Replace the Main PCA
Low Pressure failure
Leak and tubing
connections,
Compressor sieve
canister connections
Spine leaks
Faulty Product Tank
Using Real time data with a setting of
Pulse 1 in Demo mode, check the
pressure to ensure it is above 4.
Check compressor RPM and abnormal
compressor noise.
Using snoop leak detector check for
leaks in the spine.
Replace Product Tank
Hard Inter processor packet
transmission Hardware failure
Main PCA Replace the Main PCA
RAM Hardware failure at
power on
Main PCA Replace the Main PCA