PAGE 5-4
1097626, V
ER. 03
5.2 SIMPLYFLO ALARM INDICATORS AND SCREEN SYMBOLS
VISUAL, AUDIO INDICATORS DESCRIPTION WHAT TO DO
A sequence of 3 long beeps
+
1 short beep that repeats every
4 seconds
Low Oxygen
Concentration Alarm
This alarm occurs when
the device is delivering a
lower concentration of
oxygen than specified.This
alarm can be silenced by
pressing the Silence
button.
Note: This alarm occurs
when the internal O2
sensor reads <82%
O2.This alarm will turn off if
the system is able to
recover from the fault and
achieve >84% O2.
Change to another source
of oxygen and contact your
home care provider.
A sequence of 3 short beeps
+
3 long beeps that repeats every
4 seconds.
Technical Fault Alarm
This alarm occurs when
the device has a general
malfunction and the device
is no longer operating
properly.
The device may or may not
shut down depending on
the severity of the alarm.
This alarm can be silenced
if the
symbol appears on-screen.
If not, the user must turn
off the power to silence the
alarm.
If this alarm occurs again,
change to another source
of oxygen and contact your
home care provider.
A sequence of 3 long beeps
+
1 short beep that repeats every
4 seconds.
No Flow Alarm
This alarm occurs when
the device detects there is
no oxygen flowing in the
patient cannula.This alarm
can be silenced by
pressing the Silence
button.
Check the cannula for
kinks or other obstructions
that are stopping oxygen
flow through it.