PAGE 9-11097626, VER. 03
CHAPTER 9: TESTING
9.0 CHAPTER OVERVIEW
This chapter details the necessary procedures to perform testing procedures on the SimplyGo Portable
Oxygen Concentrator and the SimplyFlo Transportable Oxygen Concentrator.
9.1 SIMPLYGO FINAL TESTING PROCEDURE
9.1.1 REQUIRED EQUIPMENT
• Oxygen Analyzer (PRI P/N 1027705) or equivalent
• Pulsemeter - (Valley Inspired Pulse Meter (http://www.inspiredrc.com/pulsemeter.html, Part
Number: Pulsemeter))
• Continuous Flow meter - Cole-Parmer Hand-held Flowmeter, PM2.20XY (Catalog Number: WU-
32500-80)
• O2 Enrichment Port Adapter - (Philips Respironics Part Number: 312710)
• Test Orifice Cap - (www.caplugs.com, Part Number: EC-14) or equivalent
• Oxygen Interface Hardware Kit (Respironics Part Number: 1073499)
• Computer with Microsoft Windows 7 (32 bit) or XP (32 bit) operating systems only
• Current version of the Data Recovery Tool for SimplyGo (my.respironics.com)
• Current version of the SimplyGo Manifest File (my.respironics.com)
9.1.2 PROCEDURE
1. Verify that a battery is properly installed in the SimplyGo.
2. Connect the AC Power Supply to an AC power source.
IMPORTANT NOTE
If a Main PCA, Sieve Bed, Compressor, or Process Valve has been replaced, perform the Final Test Pro-
cedure and then the Balancing Procedure in that order.
If any other repair has been completed, perform only the Final Test Procedure.
If performing the Final Test and Balancing, use the Final Test and Balancing Data Sheet.
If performing the Final Test only, use the Final Test Data Sheet.
NOTE
For software download and installation instructions, perform all steps in Section 5.3.2 through 5.3.4 of this
manual.