PAGE 9-2
1097626, V
ER. 03
3. Connect the AC Power Supply output connector to the SimplyGo power input connector and listen
closely to verify that the cooling fan turns on. Record results.
4. Connect the Pulsemeter tubing to the patient cannula connector.
5. Launch the SimplyGo Data Recovery Tool by double clicking on the icon shown below.
6. Ensure the device is connected to the computer via the communication cable
7. Turn the device on by performing the following steps:
8. Press the Power button
9. Wait for the LCD screen to light
10. Press the Power button a second time
11. The data recovery tool should now detect the device and begin downloading the data
12. Once the data has been downloaded, proceed to the code loading option.
13. Document the device model number, PCA revision and motor type in the Test Data Sheet. These
values will be needed to select the appropriate software and parameters to load.
14. Load the appropriate software and parameters for the device configuration you have.
NOTE
The fan may continue to run if the battery requires charging but will stop if the battery does not require
charging. This is normal operation.