PAGE 5-2
1097626, V
ER. 03
3 long beeps + 1 short beep
No Flow Alarm
This alarm occurs when the device detects there is no oxygen flowing in
the patient cannula.
Alternating short and long
beeps
Depleted Battery Alarm
This alarm occurs when approximately two minutes of battery life
remains. The battery symbol flashes on-screen.
Alarm Silence Symbol
This symbol appears when you press the Alarm Silence button to
silence the audible alarm (beeps).
Attention
This symbol appears when corrective action is required.
Battery Charging While Almost Empty Symbol
This battery symbol appears (with the charging bars scrolling on the left
side) when the battery is depleted and recharging.
Battery Charging While Partly Full Symbol
When a battery symbol similar appears (with the charging bars scrolling
on the left and at mid-height on the right), the battery is partly full and
recharging.
The left side of the symbol indicates charging, and the right side
indicates the battery level.
1 long beep
Battery Charge Complete Symbol
This battery symbol appears when the battery is fully charged.
Power Connection Symbol
This symbol appears when the device is connected to an AC, DC, or
airline power source.